“Equipment specifically built to empty, wash, rinse, thermally disinfect and cool bedpans and urinals The Tornado™ system comprises of a series of high pressure water cycles delivered through strategically located fixed and rotating nozzles, followed by a steam phase.
Key features
Easy to use - simply load, close and activate programme.
A flexible design - choice of polymeric material or stainless steel for outer casing.
9+2 - system includes nine fixed and two rotating nozzles to meet the highest cleaning requirements.
Pipe System Disinfection (PSD) provides rapid disinfection through all fixed and rotating nozzles, disinfecting the entire pipe and nozzle system with every cycle.
Smooth, hygienic chamber means no folds or pockets.
Safe and rapid thermal disinfection of the goods with the steam transported into the chamber through all pipes and nozzles.
Meets and exceeds the stringent cleaning and disinfection requirements of EN ISO15883 (CAN/CSA Z15883).
Drying/cooling fan for rapid and hygienic dry and cooling of recepticles with HEPA filtered (H13) air.
Deep-drawn stainless steel chamber without welded joints.
Read out display provides the user with all information needed for operation and control.
Remaining time displayed in an easy-to-follow cycle sequence.
Ao value and control.
Three programmes available: economy, normal and intensive, with 22 additional programmes available upon request.
Optional automatic door function.
Locking system during cycle, securing staff safety and ensuring a complete cleaning and disinfection process.
Optional process records, via printer or USB.
Environmentally friendly - low consumption of water and energy, recyclable material.
Wide range of holders available for different receptacles.
Lockable storage for detergent containers.
Easy maintenance and service.“”
Tornado - TECHNICAL DATA
UTILITY REQUIREMENTS
Utility Unit connection Supply Pressure requirements Flow
requirements Comments
Cold Water (CW) 15 mm (G1/2") 100-800 kPa (1-8 bar, 14-116 PSi) 20 L/min See notes 1 and 3
Hot Water (HW) 15 mm (G1/2") 100-800 kPa (1-8 bar, 14-116 PSi) 20 L/min See notes 2 and 3
Drain (D) Ø90 alt. Ø 110mm Not Applicable Peak 1 L/sec See notes 3 and 4
Electrical (equipped with internal steam generator)
Voltage Recommended service disconnect Full load, Amps See notes 5 and 6
L1, 230 VAC, 1N+PE, 50 Hz Max. 3,0 kW, 1x Fuse C 16 A 13 A
L2, 230VAC, 3+PE, 50Hz Max. 3,75 kW, 3x Fuse C 16 A 12 A
L3, 380-400VAC, 3N+PE, 50 HzA Max. 3,75 kW, 3x Fuse C 10 9 A
L5, 240 VAC, 2+PE, 60 Hz Max .3,3 kW, 2 x Fuse C 15 A 15 A
L6, 240VAC, 3+PE, 60 Hz Max. 3,75 kW, 3x Fuse C 15 A 13 A
L9, 208 VAC, 2+PE, 60 Hz Max. 3,0 kW, 2x Fuse C 15 A 14 A
L10, 208VAC, 3+PE, 60 Hz Max. 3,75 kW, 3x Fuse C 15 A 13 A
L14, 200 VAC, 2+PE, 50 Hz Max .3,0 kW, 2x Fuse C 16 A 15 A
L15, 200VAC, 3+PE, 50 Hz Max. 3,75 kW, 3 x Fuse C 16 A 14 A
L18, 200 VAC, 2+PE, 60 Hz Max. 3,0 kW, 2x Fuse C 16 A 15 A
L19, 200VAC, 3+PE, 60 Hz Max. 3,75 kW, 3x Fuse C 16 A 15 A
L22, 230 VAC, 1N+PE, 60 Hz Max. 3,0 kW, 1x Fuse C 16 A 13 A
L24, 380-400VAC, 3N+PE, 60 Hz Max. 3,75 kW, 3x Fuse C 10 A 10 A
L28, 240 VAC, 1N+PE, 50 Hz Max. 3,3 kW, 1x Fuse C 16 A 13,6 A
L30, 415VAC, 3N+PE, 50 Hz Max. 3,75 kW, 3x Fuse C 10 A 9 A
OPERATING CONDITIONS
Room temperature 5-40°C (41-104F)
Air humidity < 80% at 31°C (87° F)
Max surface temperature 40° C (104°F)
Water consumption
Economy program 11 liter ± 10 %*
* if option cooling/drying fan is selected for the cooling phase. If internal cooling is selected, additionally 3 liters ± 1 liter will be added to the stated consumption.
Normal program 18 liter ± 10 %*
Intensive program 25 liter ± 10 %*
Chamber size and volume
Stainless steel 565x410x290, 67 liter
Noise level
Octave band(Hz)
125 250 500 1000 2000 4000 8000 LwA LwAFmax The sound power level as
linear octave band levels are
presented as A-leveled sound
power level, both LwA and
LwAFmax.
Correction (Kok) 57 61 58 55 53 48 47 61 75
Estimated LpA levels for Tornado located in a 70 m³ room, db ref 20 μPa.
Type of room Description Operator location * 3 m from washer The calculated sound power
levels are different LpA levels in
different types of rooms. With
bigger room volume the noise
level is reduced and vice versa.
Here are three examples of
expected LpA levels.
* See Note 7
Hard All walls from concrete etc no
absorbing elements. 60 60
Normal Some sound
absorbing elements
like furniture and
textiles. 56 55
Subdued Some absorbing
elements in ceiling
and textiles. 53 51
Outer dimensions
Freestanding See installation drawing in subsection
installation.
Width 450 mm (17" 11/16)
Depth 580 mm (22" 3/4)
Height 1320 mm (51" 7/8)
Weight 100 kg/220 lb
Notes
1) Cold water recommended quality = potable water accordingly to WHO guidelines, temp. <20°C (<68°F).
2) Hot water recommended quality = potable water accordingly to WHO guidelines, temp. 45-60°C (113-140°F).
3) It shall be the Customers responsibility to eliminate water hammer conditions if they occur in the service piping.
4) It shall be the Customers responsibility to provide a properly sixed and located drainage system in accordance with the national
plumbing code. Consider that other equipment is tied to drain to eliminate slow drainage or backup.
5) It shall be the Customers responsibility to complete all electrical connections using properly sized wiring in accordance with the
national electrical code.
6) It shall be the Customers responsibility to install an approved a fused disconnect switch accordingly to national code in all electrical
supply lines at the washer disinfector location.
7) LpA levels are measured to EN ISO 3747:2000, 1 meter in front of the washer and 1,5 meter above the floor level.
Systems Available
Freestanding with or without automatic door.
Accessories
Arjo bedpan with lid.
Drainage floor, Ø 3 ½ and Ø 4 1/4" (90 and 110 mm).
Drainage wall, , Ø 3 ½ and Ø 4 1/4" (90 and 110 mm).
Valves and filter.
Arjo Clean Flusher Detergent, 5L canister.
Arjo Clean Flusher Rinse, 5L canister.“” ”