“Equipment specifically built to empty, wash, rinse, thermally disinfect and cool bedpans and urinals Available in a number of models, the Ninjo™ system comprises of a series of high pressure water cycles delivered through strategically located fixed and rotating nozzles, followed by a steam phase.
Key features
Easy to use - simply load, close and activate programme.
9+2 - system includes nine fixed and two rotating nozzles to meet the highest cleaning requirements.
Pipe System Disinfection (PSD) provides rapid disinfection through all fixed and rotating nozzles, disinfecting the entire pipe and nozzle system with every cycle.
Smooth, hygienic chamber means no folds or pockets.
Ao value display - disinfection levels displayed, confirming that the appropriate level of disinfection has been reached.
Remaining time displayed in an easy-to-follow cycle sequence.
Two programmes available: economy and normal, with an additional three available upon request.
Locking system during cycle, securing staff safety and ensuring a complete cleaning and disinfection process.
Optional drying/cooling fan for rapid and hygienic dry and cooling of recepticles with HEPA filtered (H13) air.
Meets and exceeds the stringent cleaning and disinfection requirements of EN ISO15883 (CAN/CSA Z15883).
Choice of polymeric material or stainless steel for chamber or outer casing.
Environmentally friendly – low consumption of water and energy, recyclable material.
Wide range of holders for different receptacles.
Lockable storage for detergent containers.
Easy maintenance and service.
NINJO - TECHNICAL DATA
UTILITY REQUIREMENTS
Utility Unit connection Supply Pressure requirments Flow
requirement Comments
Cold water (CW) 15 mm (G½ “) 100-800 kPa (1-8 bar, 14-116 PSi) 20 L/min See Notes 1 and 3
Hot water (HW) 15 mm (G½ “) 100-800 kPa (1-8 bar, 14-116 PSi) 20 L/min See Notes 2 and 3
Drain (D) Ø 90 mm alt.
Ø 110 mm N/A Peak 1 L/sec See Notes 3 and 4
Electrical
Voltage Recommended service disconnect Full load, Amps See notes 5 and 6
L1, 230 VAC, 1N+PE, 50 Hz Max. 3,0 kW, 1 x Fuse C 16 A 13 A
L5, 240 VAC, 2+PE, 60 Hz Max .3,3 kW, 2 x Fuse C 15 A 13,6 A
L9, 208 VAC, 2+PE, 60 Hz Max. 3,0 kW, 2x Fuse C 15 A 14,4 A
L14, 200 VAC, 2+PE, 50 Hz Max .3,0 kW, 2x Fuse C 16 A 15 A
L18, 200 VAC, 2+PE, 60 Hz Max. 3,0 kW, 2x Fuse C 16 A 15 A
L22, 230 VAC, 1N+PE, 60 Hz Max. 3,0 kW, 1x Fuse C 16 A 13 A
L28, 240 VAC, 1N+PE, 50 Hz Max. 3,3 kW, 1x Fuse C 16 A 13,6 A
OPERATING CONDITIONS
Room temperature 5-40°C (41-104F)
Air humidity < 80% at 31°C (87° F)
Max surface temperature 40° C (104°F)
Noise level
Octave band (Hz) 125 250 500 1000 2000 4000 8000 LwA LwAFmax The sound power level as linear
octave band levels are presented
as A-leveled sound power level,
both LwA and LwAFmax.
Correction (Kok) 57 61 58 55 53 48 47 61 75
Estimated LpA levels for Ninjo located in a 70 m³ room, db ref 20 μPa.
Type of room Description Operator location * 3 m from washer The calculated sound power levels
are different LpA levels in different
types of rooms. With bigger room
volume the noise level is reduced
and vice versa. Here are three
examples of expected LpA levels.
* See Note 7
Hard All walls from concrete etc no absorbing elements.
60 60
Normal Some sound absorbing elements like furniture and textiles.
56 55
Subdeued Some absorbing elements in ceiling and textiles.
53 51
Water consumption
Economy program 11 liter ± 10 %*
* if option cooling/drying fan is selected for the cooling phase. If internal cooling is
selected, additionally 3 liters
± 1 liter will be added to the stated consumption.
Normal program 18 liter ± 10 %*
Chamber size and volume
Stainless steel 565x410x290, 67 liter Polymeric 535x400x300, 64 liter
Outer dimensions
Freestanding Under counter
Width 450 mm (17" 11/16) 450 mm (17" 11/16) See installation drawing in subsection installation.
Depth 580 mm (22" 3/4) 580 mm (22" 3/4)
Height 1320 mm (51" 7/8) 870 mm (34" 3/16)
900 mm (35" 7/16) with lid
Weight 90 kg/200 lb 70 kg/155 lb
Notes
1) Cold water recommended quality = potable water accordingly to WHO guidelines, temp. <20°C (<68°F).
2) Hot water recommended quality = potable water accordingly to WHO guidelines, temp. 45-60°C (113-140°F).
3) It shall be the Customers responsibility to eliminate water hammer conditions if they occur in the service piping.
4) It shall be the Customers responsibility to provide a properly sixed and located drainage system in accordance with the national
plumbing code. Consider that other equipment is tied to drain to eliminate slow drainage or backup.
5) It shall be the Customers responsibility to complete all electrical connections using properly sized wiring in accordance with the
national electrical code.
6) It shall be the Customers responsibility to install an approved and fused disconnect switch accordingly to national code in all
electrical supply lines at the washer disinfector location.
7) LpA levels are measured to EN ISO 3747:2000, 1 meter in front of the washer and 1,5 meter above the floor level.
Systems available
Freestanding
Under-counter
Accessories
Arjo bedpan with lid
Drainage floor Ø 90 mm and 110 mm (3 1/2"” and Ø 4 1/4")
Drainage wall Ø 90 mm and 110 mm (3 1/2"” and 4 1/4")
Valves and filter
Arjo Clean Flusher™ Detergent, 5L canister
Arjo Clean Flusher™ Rinse, 5L canister”