Le connecteur AGSS est un appareil indispensable pour les blocs opératoires. Lors d'interventions médicales, l'utilisation de gaz anesthésiants contribue à polluer le bloc opératoire. Pour une opération réussie, il est important d'éliminer ces gaz anesthésiques. Le système d'évacuation des gaz anesthésiques (AGSS) crée une perte de charge et aspire les gaz polluants.
Spécification
- Intuitive use
--- Pressure reduction via the Venturi effect, with direct connection to the air network
--- Display of On/Off status via a green LED
--- Optimum adaptability with 3 types of mounting
- Safety
--- Suction stops immediately on disconnection
--- No risk of gas backflow thanks to non-return valve
--- No risk of network contamination, the AGSS fitting has its own evacuation network
- Anaesthesia under control
--- Suction stops immediately on disconnection
--- Secure suction, non-return valve and visual On/Off signal
--- No risk of network contamination, the AGSS fitting has its own evacuation network
Specifications
- Suction flow rate 45-50 L/min
- Air consumption 40 to 60 L/min
- Max. suction volume 80 L/min
- Air connection Copper pipe 6 x 8 mm
- Gas exhaust connection Copper pipe 13 x 15 mm
- Air operating pressure 5 bar
- Mounting types Trunking, flush, surface
- Dimensions of front panel 160 x 160 mm
- Dimensions of flush-mounted box 150 x 150 x 60 mm
- Plastic adaptor 20 mm
- Weight Trunking: 670 g, flush: 1850 g, surface: 2100 g
- Regulatory compliance Class IIa medical device with CE 0044 markings, distributed by Air liquide Medical Systems S.A. Read the user manual carefully. ISO 9170-2 standard.
- Life time 15 years