- Optimum safety
--- Safety plate: continuous supply, a network shut-off is impossible
--- 2 pressure gauges: 1 upstream and 1 downstream, colour-coded to facilitate pressure control
--- The cover protects against chock and dust
--- Quick and easy action in case of emergency
- Unrivalled simplicity
--- On/Off in just one action
--- Compatible with all gases
--- Easy and foolproof reversal of gas fl ow direction
--- Optimised space: compact, lightweight with simplifi ed installation
- Easier maintenance
--- Control lever for fast action in a single motion Interchangeable pressure reduction module, positioned vertically, no risk of falling
--- Continuous gas distribution because modules can be maintained separately
- High performance for gas network
--- Intuitive use: 3 gas fl ow confi gurations for continuous supply to the network
--- Maintenance is facilitated by the control levers
--- Safety plate: prevents forgetting and mistakes
Specifi cations
- Available gases O2, Air, N2O, CO2, N2, Air-800
- Gas inlet standards NF and on request DIN, BS, UNI
- Supply pressure 5.5 to 10 bar
- Downstream pressure Preset to 4 bar (8 bar for Air-800), adjustable from 0 to 10 bar
- Maximum fl ow rate 40 Nm3/h
- Network connection Ready-to-braze copper tube, diameter 10x12 mm
- Sensor connection M10 x 100
- Weight of double pressure reducer 3,95 kg (cover not included)
- Weight of single pressure reducer 2,91 kg (cover not included)
- Mass of cover 0,3 kg
- Mass of magnetic materials Single Damao: 80 g / Double Damao: 145 g
- Dimensions of double pressure reducer 157,5 mm (H) x 259 mm (L) x 107 mm (l)
- Dimensions of cover 216 mm (H) x 303 mm (L) x 120 mm (l)
- Upstream pressure gauge 0 to 16 bar
- Downstream pressure gauge 0 to 10 bar (0 to 16 bar for the Air-800)
- Pressure gauge precision ± 2.5% of full scale
- Regulatory compliance Class IIb medical device with CE 0459 marking, manufactured by Air Liquide Medical Systems S.A. ISO 10524-2 / ISO 7396-1 standards. Read the user manual carefully.
- Life time 10 years