Ventilkästen sind wesentliche Komponenten für die Erstellung von Versorgungsverteilern für die Hochdrucktankstelle. Dieses Gerät, kombiniert mit einem Pigtail, wird verwendet, um die Flasche mit der Tankstelle zu verbinden.
Spezifikation
- Modularity
--- Choice of confi guration, 1 or 2 cylinders connected
--- Kit assembly, possibility of connecting several valve-boxes
--- Adaptable to a main or backup manifold
- High level of safety
--- Automatic regulation of the manifold pressure thanks to its controlled pressure reduction system
--- No risk of pigtail fl apping when changing the cylinders thanks to its non return valve
--- Limited risk of adiabatic compression thanks to its “zero plastic” design
- Performance
--- Adapts perfectly to all medical gases
--- Equalisation of manifold pressure, for a real visualisation of the pressure on the gauges
--- Facilitated leak detection; the measured pressure matches the overall pressure of the manifold
-Controlling high pressure
--- Adapts perfectly to all medical gases
--- Ideal confi guration, kit assembly
--- Pressure equalisation and self-regulation
Specifi cations
- Operating pressure Up to 200 bar
- Type of fi tting 1/4” gas HP fi tting for copper tube 5x10 mm
- Dimensions 67 mm (H) x 112 mm (L) x 26 mm (l)
- Weight 1 kg
- Material Brass
- Wall mounting 2 holes ø 8 mm, spaced at 90 mm, with brackets and screws provided
- Regulatory compliance Class IIa medical device with CE 0459 marking, manufactured by Air Liquide Medical Systems S.A. ISO 7396-1 standard. Read the user manual carefully.
- Life time 15 years